• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY WOLVERINE¿ CORONARY CUTTING BALLOON® MONORAIL®; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY WOLVERINE¿ CORONARY CUTTING BALLOON® MONORAIL®; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number H74939401153000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 04/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age at time of event: approximately 65-70 years.Device evaluated by mfr: the device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is anticipated procedural complication as the event is due to a known physiological effect of the procedure noted within the directions for use, and/or device labeling.(b)(4).
 
Event or Problem Description
It was reported that a vessel dissection occurred.The 80% stenosed target lesion was located in a mildy tortuous mid right coronary artery.A 15mm x 3.00mm wolverine¿ coronary cutting balloon® monorail® was inflated and caused a moderate dissection in the artery.The device was removed and a regular balloon catheter was used, then the dissection was stented with a non-bsc stent.There were no further patient complications and the patient status is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE¿ CORONARY CUTTING BALLOON® MONORAIL®
Common Device Name
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7533162
Report Number2134265-2018-04170
Device Sequence Number18862880
Product Code NWX
UDI-Device Identifier08714729888352
UDI-Public08714729888352
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P950020/S072
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/06/2019
Device Model NumberH74939401153000
Device Catalogue Number39401-15300
Device Lot Number0021478315
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2018
Initial Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-