| Model Number |
H74939401153000 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Intimal Dissection (1333)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: approximately 65-70 years.Device evaluated by mfr: the device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is anticipated procedural complication as the event is due to a known physiological effect of the procedure noted within the directions for use, and/or device labeling.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a vessel dissection occurred.The 80% stenosed target lesion was located in a mildy tortuous mid right coronary artery.A 15mm x 3.00mm wolverine¿ coronary cutting balloon® monorail® was inflated and caused a moderate dissection in the artery.The device was removed and a regular balloon catheter was used, then the dissection was stented with a non-bsc stent.There were no further patient complications and the patient status is fine.
|
| |
|