| Model Number |
GL-62-MG |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during procedure, when trying to detach the thread, the needle broke.The procedure was completed with another device.
|
| |
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.The visual inspection of the opened samples noted three sutures and four needles.One needle was received bent and broken at the swaged end.Upon microscopic inspection, instrument marks were observed near the break site.As no sealed samples were returned, a bend moment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.The needle should always be held in the middle where the diameter is greatest.Grasping the needle near the swaged end or the tip could cause bends or breaks, or damage the connection between the needle and suture.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|