• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. STRATAFIX SPIRAL PDS+; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. STRATAFIX SPIRAL PDS+; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE Back to Search Results
Catalog Number SXPP1B420
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that a patient underwent a reconstruction operation using flap on an unknown date and a barbed suture was used.During continuous suturing on the fascia, the surgeon felt the barbs on the suture were not caught on the tissue firmly.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Date sent to the fda: (b)(4).A dispensed needle-suture combination and one empty labeled d winding was received for analysis.During the visual inspection of the sample, the swage and attachment area were as expected.The suture was examined and the barbs were present along of the strand; also, ten barbs were measured and the barbs were barbs too shallow less of the minimal depth, resulting in a smaller barb.Per the sample condition, the assignable cause of the performance barb non-engagement in tissue is barb depth out of specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX SPIRAL PDS+
Type of Device
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
Manufacturer (Section G)
ETHICON, INC. SAN LORENZO
road 183, km. 8.3
san lorenzo
Manufacturer Contact
darlene kyle
p.o. box 151, route 22 west
somerville, NJ 08876-0151
9082182792
MDR Report Key7534479
MDR Text Key108989750
Report Number2210968-2018-72971
Device Sequence Number1
Product Code NEW
UDI-Device Identifier10705031235809
UDI-Public10705031235809
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K150670
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSXPP1B420
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received06/04/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-