PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C0530
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2018-01032.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, the physician successfully deployed and detached a ruby coil in the target vessel using a non-penumbra microcatheter.While attempting to advance a new ruby coil through the microcatheter, the physician experienced resistance and the ruby coil would not advance; therefore, it was removed.The physician then successfully deployed and detached another ruby coil in the target vessel using the same microcatheter.While attempting to advance a new ruby coil, the physician experienced resistance again and the ruby coil would not advance; therefore, it was removed.The procedure was completed using a new ruby coil and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the device in complaint was expected to be returned; however, additional information received from the penumbra sales representative indicated that the device was disposed of and is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7534655
Report Number3005168196-2018-01033
Device Sequence Number1551093
Product Code HCG
UDI-Device Identifier00814548012919
UDI-Public00814548012919
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/02/2025
Device Catalogue NumberRBY2C0530
Device Lot NumberF79807
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/26/2018
Supplement Date Received by Manufacturer06/21/2018
Initial Report FDA Received Date05/22/2018
Supplement Report FDA Received Date06/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
-
-