• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DUTCH OPTHALMIC RESEARCH CENTER BV EVA, COMBINED MACHINE INCLUDING LASER (SMA CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DUTCH OPTHALMIC RESEARCH CENTER BV EVA, COMBINED MACHINE INCLUDING LASER (SMA CONNECTOR) Back to Search Results
Model Number 8000.COM02
Medical Device Problem Code Device Operational Issue (2914)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/05/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump module has been investigated without conclusive result, and reworked to specification.You have already received replacement for this complaint.Based upon these facts the complaint has been closed.Considering the failure, pum152 (temperature error: motor broken.) error showed up which requires replacement of motor.Hence can be concluded as reportable.(b)(4).No information reasonable suggests anyone was harmed.All available information has been disclosed.This complaint has now been closed.
 
Event or Problem Description
Pump error 152 - high temperature = motor failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVA, COMBINED MACHINE INCLUDING LASER (SMA CONNECTOR)
Common Device Name
EVA, COMBINED MACHINE INCLUDING LASER (SMA CONNECTOR)
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7535386
Report Number1222074-2018-00147
Device Sequence Number1121824
Product Code HQC
UDI-Device Identifier08717872019659
UDI-Public08717872019659
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 02/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number8000.COM02
Device Catalogue Number8000.COM02
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2016
Initial Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-