| Model Number |
FEM12040 |
| Medical Device Problem Codes |
Material Perforation (2205); Misfire (2532)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was returned to the manufacturer for evaluation.The investigation is currently under way.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure in a left arm av fistula to treat a venous stenosis, the stent graft allegedly partially deployed.It was further reported that an alleged strut perforation on the catheter was identified under fluoroscopy; therefore, the catheter was removed without incident from the patient.Reportedly, angioplasty was performed with a pta balloon resulting in vascular flow and completion of the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: based on the available information and the investigation performed, there is no indication that the product failed to meet its specification prior to shipment.The lot records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.The review shows that no remarkable incidents occurred during the manufacturing process.No relevant manufacturing process changes were implemented, that could have led to the event reported.Investigation summary: on the basis of the returned stent graft delivery system, the alleged deployment failure could be confirmed.The stent graft was found partially released and the outer sheath was found perforated by stent graft strut, which made a stent graft deployment impossible.No indication was found for manufacturing related issues.As a result of the investigation performed the complaint is confirmed.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risk.The ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend including the terminal cephalic arch or across the elbow joint has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." regarding the use of accessories the ifu states: "materials required for the fluency plus endovascular stent graft procedure: introducer sheath with appropriate inner diameter".Furthermore, the ifu states: "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment.".
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure in a left arm av fistula to treat a venous stenosis, the stent graft allegedly partially deployed.It was further reported that an alleged strut perforation on the catheter was identified under fluoroscopy; therefore, the catheter was removed without incident from the patient.Reportedly, angioplasty was performed with a pta balloon resulting in vascular flow and completion of the procedure.There was no reported patient injury.
|
| |
|