Model Number ZXT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the toric intraocular lens (model zxt150 +17.0 diopter) was implanted in the right eye of a female patient on (b)(6) 2017.The lens was explanted on (b)(6) 2018 and replaced with a competitor non-toric lens.The reason for the explant was not provided.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 5/29/2018.Device returned to manufacturer? yes.Device evaluation: the sample was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the device were reviewed.The product was manufactured and released according to specification.Historical data analysis: the search in complaint history revealed no similar complaints were received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7535501
Report Number9614546-2018-00492
Device Sequence Number18890303
Product Code POE
UDI-Device Identifier05050474605145
UDI-Public(01)05050474605145(17)220424
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date04/24/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/24/2018
Supplement Date Received by Manufacturer08/10/2018
Initial Report FDA Received Date05/22/2018
Supplement Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-