| Model Number |
0601620 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the facility has alleged the catheter has holes in the tubing, small splits in the tubing and/or separation at the repair kit.Device was removed.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Investigation summary: one white hub hickman extension leg was returned for evaluation.Visual and microscopic inspection identified a partial circumferential break just distal to the white hub.Therefore, the investigation is confirmed for the alleged leak.The definitive root cause could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter allegedly had holes in the tubing, small splits in the tubing, and/or separation in the repair kit.It was further reported that the device was removed.There was no patient injury.
|
| |
|