The device was not returned for evaluation.Imaging was received and reviewed by william cook australia medical director, who stated "the images fit the story provided in the complaint summary.The leak was from the right renal fenestration below the renal bridging stent.This would make it a type iii endoleak, and not a type i as described in the complaint form." work order ac949549 was reviewed and appears complete and correct.The device ifu states: "the long-term performance of fenestrated endovascular grafts, including the stents placed in fenestrations/scallops, has not yet been established.All patients should be advised that endovascular treatment requires life-long, regular follow-up to assess their health and the performance of their endovascular graft.Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysms, changes in the structure or position of the endovascular graft, or stenosis/occlusion of vessels accommodated by fenestrations) should receive enhanced follow-up." "after endovascular graft placement, patients should be regularly monitored for perigraft flow, aneurysm growth, patency of vessels accommodated by a fenestration/scallop, or changes in the structure or position of the endovascular graft.At a minimum, annual imaging is recommended, including: abdominal radiographs to examine device integrity (separation between components, stent fracture or barb separation) contrast and non-contrast ct to examine aneurysm changes, perigraft flow, patency, tortuosity and progressive disease.If renal complications or other factors preclude the use of image contrast media, abdominal radiographs and duplex ultrasound may provide similar information." "intervention or conversion to standard open surgical repair following initial endovascular repair should be considered for patients experiencing enlarging aneurysms, unacceptable decrease in fixation length (vessel and component overlap) and/or endoleak.An increase in aneurysm size and/or persistent endoleak may lead to aneurysm rupture." "potential adverse events that may occur and/or require intervention include, but are not limited to: endoleak; endoprosthesis: improper component placement; incomplete component deployment; component migration; suture break; occlusion; infection; stent fracture; graft material wear; dilatation; erosion; puncture; perigraft flow; barb separation and corrosion." "all patients should be advised that endovascular treatment with this device requires lifelong, regular follow-up to assess their health and the performance of their endovascular graft.Patients with specific clinical findings (e.G., endoleaks, enlarging aneurysms, or changes in the structure or position of the endovascular graft) may require additional follow-up.Patients should be counseled on the importance of adhering to the follow-up schedule, both during the first year and at yearly intervals thereafter.Patients should be informed that regular and consistent follow-up is a critical part of ensuring the ongoing safety and effectiveness of endovascular treatment of abdominal aneurysms with this device.Physicians should evaluate patients on an individual basis and prescribe follow-up relative to the needs and circumstances of each individual patient." from the information received in this complaint, a definitive root cause could not be determined to identify the cause of the type iii endoleak at the right renal fenestration below the renal bridging stent.It is possible that one or more of the following factors contributed the complaint: insufficient sizing of the renal bridging stent.Insufficient flaring of the proximal end of the renal stent.Insufficient positioning of renal bridging stent.Factors relating to the procedure.
|