AV-TEMECULA-CT RX HERCULINK ELITE STENT SYSTEM; RENAL STENT SYSTEM Back to Search Results
Catalog Number 1011493-18
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The instructions for use states: the rx herculink elite renal stent system is indicated for use in patients with atherosclerotic disease of the renal arteries following sub-optimal percutaneous transluminal renal angioplasty (ptra) of a de novo or restenotic atherosclerotic lesion (15 mm in length) located within 10 mm of the renal ostium and with a reference vessel diameter of 4.0 and 7.0 mm.The investigation determined that the reported difficulties were due to case circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified iliac artery.As a 5.0 x 18mm herculink elite stent delivery system was being advanced, the physician deemed it necessary to use a different size device.During removal, the delivery system got stuck on the guiding catheter (gc).The device was removed as a single unit including the gc.Another herculink was used to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Common Device Name
RENAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7535748
Report Number2024168-2018-03833
Device Sequence Number16619148
Product Code NIN
UDI-Device Identifier08717648078026
UDI-Public08717648078026
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P110001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number1011493-18
Device Lot Number7071961
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2018
Initial Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient Weight112
-
-