• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call the insulin pump alarmed motor error and a motor position encoder error.The customer¿s blood glucose level was unknown at the time of the incident.The customer stated that the drive support cap was normal.The customer stated that the insulin pump was not exposed to high magnetic fields.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the self-test, unexpected restart error test, displacement test, rewind, basic occlusion test, occlusion test, prime test and excessive no delivery test or verify operating currents, motor position encoder error, motor position encoder error and motion sensor test failure alarm due to motor error alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7536274
Report Number3004209178-2018-73444
Device Sequence Number16619000
Product Code OYC
UDI-Device Identifier00643169932951
UDI-Public(01)00643169932951
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAH
Device Catalogue NumberMMT-523LNAH
Device Lot NumberA5523LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2018
Initial Date Received by Manufacturer 04/26/2018
Supplement Date Received by Manufacturer08/28/2018
Initial Report FDA Received Date05/23/2018
Supplement Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight135
-
-