Model Number 71812
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheters were kinked.The user stated that one box of 15 had kinks in them and could not be used.Another box of 22 also had kinks.The patient stated that she did not use the catheters once the packet was opened.
 
Event or Problem Description
It was reported that the catheters were kinked.The user stated that one box of 15 had kinks in them and could not be used.Another box of 22 also had kinks.The patient stated that she did not use the catheters once the packet was opened.
 
Additional Manufacturer Narrative
The reported event was confirmed; however, the cause is unknown.Per visual inspection, the samples were bend near the mid-section.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: " review the self-catheterization procedure with a healthcare professional.1.Always wash hands prior to use.2.Open the catheter package by peeling the tab upwards with the aid of the finger hole.3.If desired, hang the package with the adhesive surface on the inner side of the tab to a nearby dry vertical surface while preparing to catheterize.4.Positioned comfortably with thighs spread apart, clean the opening to the urethra ¿ and the surrounding area.Wipe from front to back to avoid fecal contamination.Wash your hands again.5.Remove the catheter with the aid of the insertion handle and advance the catheter tip into the urethra.Slowly and gently insert the catheter into the urethra until urine begins to flow (approximately 1-1.5" or 2.5-3.8 cm).6.When urine stops flowing, begin to withdraw the catheter.It is recommended to slowly rotate the catheter during withdrawal, stopping each time urine begins to flow.Check the color, odor and clarity of the urine.Any changes may need to be reported to a health care professional.Disposal 7.Place the catheter back into the package and dispose in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Do not flush down the toilet.8.Wash hands." correction: mfr site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROSIL GO¿ INTERMITTENT URINARY CATHETER
Common Device Name
INTERMITTENT CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7536494
Report Number1018233-2018-01803
Device Sequence Number13799337
Product Code EZD
UDI-Device Identifier00801741097997
UDI-Public(01)00801741097997
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K150345
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/28/2019
Device Model Number71812
Device Catalogue Number71812
Device Lot NumberJUAY0689
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2018
Initial Date Received by Manufacturer 04/30/2018
Supplement Date Received by Manufacturer09/04/2018
Initial Report FDA Received Date05/23/2018
Supplement Report FDA Received Date09/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No