Model Number W1TR01
Device Problems Mechanical Problem (1384); Misconnection (1399)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/21/2018
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.The returned device indicated a grommet issue.The returned device indicated a set screw with a rounded socket.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during an implant procedure, the left ventricular port of the device failed to allow the lead pin to fully seat.The physician had difficulty inserting and securing the lead pin and could not visualize insertion.A discoloration of the device header was also reported.The device was not used and another was implanted.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPTA CRTP
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7537386
MDR Text Key109047792
Report Number3004209178-2018-11642
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169735637
UDI-Public00643169735637
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2019
Device Model NumberW1TR01
Device Catalogue NumberW1TR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2018
Initial Date FDA Received05/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4965 LEAD
Patient Age4 YR
-
-