• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD CARIBE LTD. BD MICROTAINER® SST¿ - AMBER; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD CARIBE LTD. BD MICROTAINER® SST¿ - AMBER; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 365978
Medical Device Problem Code Incorrect Or Inadequate Test Results (2456)
Health Effect - Clinical Code No Information (3190)
Date of Event 11/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Retention samples were selected from bd inventory for evaluation, and the customer's indicated failure mode was not observed as all samples met specifications.A review of the manufacturing records was completed for the incident lot and no issues were identified.Bd has initiated a capa to document further investigation and root cause(s) of this product issue.
 
Event or Problem Description
It was reported while using bd microtainer® sst¿ - amber, that the gel would not move to separate the serum while being centrifuged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® SST¿ - AMBER
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
MDR Report Key7537443
Report Number2618282-2018-00397
Device Sequence Number11390999
Product Code JKA
UDI-Device Identifier50382903659789
UDI-Public50382903659789
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K991702
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2018
Device Catalogue Number365978
Device Lot Number7093873
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2017
Initial Report FDA Received Date05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-