Model Number 45-750812
Medical Device Problem Codes Detachment Of Device Component (1104); Separation Failure (2547)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history record indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The coil remains implanted in the patient and the remainder of the device was discarded at the user facility; therefore, a product evaluation could not be performed.The root cause is unknown.
 
Event or Problem Description
It was reported that during placement of an embolization coil, the coil did not detach.During removal, the coil detached, and was left implanted in the patient.There was no reported intervention or patient injury.The patient was reported to be doing fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR CX DETACHABLE
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key7537775
Report Number2032493-2018-00092
Device Sequence Number13907808
Product Code KRD
UDI-Device Identifier00812636021058
UDI-Public(01)00812636021058(11)160407(17)210407(10)16040711
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K151358
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/07/2021
Device Model Number45-750812
Device Lot Number16040711
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2018
Initial Report FDA Received Date05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-