STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5530-G-411
Medical Device Problem Codes Material Deformation (2976); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 03/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the reported lot.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
 
Event or Problem Description
Patient needed knee revision.The reason is ¿instability¿ in the knee.
 
Additional Manufacturer Narrative
Product available to stryker; returned to manufacturer on.An event regarding revision due to instability involving a triathlon insert was reported.The event was confirmed through clinician review of the medical records provided.Method & results: device evaluation and results:material analysis report (mar) indicated: the part was examined with the aid of a stereo microscope at magnifications up to 50x.Damage consistent with the explantation process and contact against the femoral component were observed on the insert.Impression markings were observed on the distal surface of the insert, consistent with contact against the baseplate.Burnishing, scratching and third-body indentations were observed on the insert.These are common damage modes of uhmwpe.Explantation damage and markings consistent with contact against the baseplate were also observed on the insert.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: the medical review indicated: baseplate malposition in excessive posterior tibial slope with choice for cr type of components in a knee with long standing stability problems has contributed to early failure of the triathlon cr arthroplasty with recurrent instability requiring exchange to triathlon ts.Device history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the mar indicated that the part was examined with the aid of a stereo microscope at magnifications up to 50x.Damage consistent with the explantation process and contact against the femoral component were observed on the insert.Impression markings were observed on the distal surface of the insert, consistent with contact against the baseplate.Burnishing, scratching and third-body indentations were observed on the insert.These are common damage modes of uhmwpe.Explantation damage and markings consistent with contact against the baseplate were also observed on the insert.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.The medical review indicated that baseplate malposition in excessive posterior tibial slope with choice for cr type of components in a knee with long standing stability problems has contributed to early failure of the triathlon cr arthroplasty with recurrent instability requiring exchange to triathlon ts.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient needed knee revision.The reason is ¿instability¿ in the knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7537882
Report Number0002249697-2018-01553
Device Sequence Number1775706
Product Code MBH
UDI-Device Identifier07613327050486
UDI-Public07613327050486
Combination Product (Y/N)N
Initial Reporter CountrySE
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Catalogue Number5530-G-411
Device Lot NumberJJ0MKM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2018
Initial Date Received by Manufacturer 04/26/2018
Supplement Date Received by Manufacturer07/16/2018
Initial Report FDA Received Date05/23/2018
Supplement Report FDA Received Date08/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age59 YR
Patient Weight66
-
-