U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 4MM X 12CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 4MM X 12CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0036124120
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device is not available.
 
Event or Problem Description
It was reported that during coil embolization procedure, the physician crimped the tip of the sheath strongly into the hub of the microcatheter causing the subject coil distal tip to break.Another coil was used to complete the procedure.No clinical consequences reported to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, physical as well as functional testing could not be performed.Based on the additional information provided it is most likely that the main coil broke due to a lack of continuous flush during the procedure.Lack of continuous flush can result in friction issues causing the coil to break.As per the device directions for use (dfu): "in order to achieve optimal performance of the target detachable coil system and to reduce the risk of thromboembolic complications, it is critical that a continuous infusion of appropriate flush solution be maintained between the femoral sheath and guiding catheter, the 2-tip microcatheter and guiding catheters, and the 2-tip microcatheter and stryker neurovascular guidewire and delivery wire.Continuous flush also reduces the potential for thrombus formation on, and crystallization of infusate around, the detachment zone of the target detachable coil".Because the device dfu contains specific instructions regarding use of continuous flush, which was omitted, an assignable cause of use error was assigned for the reported event.In addition, it was reported that the top of the sheath was strongly crimped into the hub of the microcatheter, and also, a non-stryker microcatheter was used during this procedure which may also contributed to the reported failure due to potential incompatibility of a non-stryker device due to the potential incompatibility of non-stryker neurovascular devices with the target detachable coil system, the use of other manufacturer¿s device(s) with the target detachable coil system is not recommended as per dfu.
 
Event or Problem Description
It was reported that during coil embolization procedure, the physician crimped the tip of the sheath strongly into the hub of the microcatheter causing the subject coil distal tip to break.Another coil was used to complete the procedure.No clinical consequences reported to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 4MM X 12CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7539149
Report Number3008881809-2018-00239
Device Sequence Number13897075
Product Code HCG
UDI-Device Identifier07613327128338
UDI-Public07613327128338
Combination Product (Y/N)N
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Catalogue NumberM0036124120
Device Lot Number19704796
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/01/2018
Supplement Date Received by Manufacturer05/27/2018
Initial Report FDA Received Date05/23/2018
Supplement Report FDA Received Date06/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-