• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM, 2D; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM, 2D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM, 2D; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM, 2D Back to Search Results
Model Number SDM-00001-2D
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 04/25/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the c-arm rotated uncommanded, past the normal detents.No injury was reported.It was determined that the circuit board needed to have the detents activated.Once this was completed the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM, 2D
Common Device Name
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM, 2D
Manufacturer (Section D)
HOLOGIC, INC.
36 & 37 apple ridge road
danbury CT 06810
MDR Report Key7540004
Report Number1220984-2018-00100
Device Sequence Number1818093
Product Code MUE
UDI-Device Identifier15420045505582
UDI-Public15420045505582
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P010025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Radiologic Technologist
Type of Report Initial
Report Date (Section B) 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Radiologic Technologist
Device Model NumberSDM-00001-2D
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2018
Initial Report FDA Received Date05/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-