| Catalog Number |
01.26.45.0048 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lot 176622: (b)(4) items manufactured and released on 12-february-2018.Expiration date: 2023-02-01.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Instrument involved: amis cup impactor terminal-m10 reference (b)(4).Lot 1755174: (b)(4) items manufactured and released 25-january-2018.No anomalies found related to the issue.To date, no similar event on the same lot has been registered.
|
| |
|
Event or Problem Description
|
|
The cup loosened from the terminal during the implant repositioning.The surgeon noticed that the cup thread was damaged.A new cup was implanted successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection performed by r&d product manager on 23 july 2018.The analysis has been conducted both with the pieces available and images attached to this complaint.During the analysis it is evaluated that the threaded part of the terminal is scratched.In addition also the threaded part of the implant hole is broken.It seem that the user has not completed the screwing of the terminal inside the cup before proceeding with the impaction phase and this caused extra stress on the threaded part causing the fracture.
|
| |
|