| Model Number |
55-0536 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Not Applicable (3189); No Code Available (3191)
|
| Date of Event |
04/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Der states patient had srom metal on metal (b)(6) 2008.He was revised bilaterally on (b)(6) 2014 with a biolox delta 36+0 head and a 54+41 altrx liner.The biolox on the right hip shattered causing the trunion to penetrate the liner which it did portrude and articulate on the inside of the metal shell.The trunion was deformed forcing a well fixed stem to be removed.The femur was shattered in the process.It was also stated that the liner was worn due to trunion.Doi: (b)(6) 2008 for stem and sleeve; doi- (b)(6) 2014 head and liner; dor: (b)(6) 2018; right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint:(b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
|
| |
|
Event or Problem Description
|
|
Ppf alleges fracture component after review of medical record the patient was revised to address failed right thr and broken femoral head.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.No code available (3191) is used to capture surgical intervention and surgery porlonged.
|
| |
|