Patient id ¿ case # also reported as (b)(6).This information is unknown.Device is an instrument and is not implanted/explanted.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.A device history records review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that during an orthopedic procedure on (b)(6) 2018, the surgeon bent the pointed tips of a set of reduction forceps with points broad-ratchet on purpose to accommodate what he needed for the surgery, but in the process, the surgeon bent the tips too much.Fragments were generated from the broken device but were easily retrieved.The surgeon used another synthes forceps to finish the surgery.There were no surgical delays reported.The procedure was successfully completed with the patient in stable/good condition.This report is for a reduction forceps with points broad-ratchet.This is report 1 of 1 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part 398.41, lot 5708806/a7ra04: release to warehouse date: january 28, 2008.Manufacture site: tuttlingen.A review of the device history records (dhr) showed that there were no issues at the time of manufacturing of this device and its sub components that would contribute to the complaint condition.No non-conformance reports (ncrs) were generated during the production of this device.A product investigation was completed: visual inspection performed at customer quality observed that the given forceps was returned with one of its tips broken at the portion proximal to its pivoting screw.The broken fragments observed to be generated during breakage were not returned.The other tip was observed with a dent at its extreme tip portion.No new issues were identified on the remaining portions of the device.The complaint condition agree with the complaint description and the complaint was confirmed.Relevant drawings for the returned device were reviewed.Dimensional inspection performed proximal to breakage measured the thickness of the handle shaft and was within the specifications.Review of the device history record(s) showed that there were no issues including during the manufacture of the product that would contribute to this complaint condition.Also based of the age of the device (10+ years) and the reported use error in complaint condition there is no indication that the material and its properties could have contributed to the complaint condition.Also, it was stated in complaint description that the surgeon tried to bend the tips in surgery to accommodate what he requires which in turn resulted in tip breakage due to intended additional forces encountered by the device.Therefore, it is confirmed that use error resulted in this broken complaint condition.During the investigation, no product design issues or discrepancies were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|