BECTON DICKINSON MEDICAL SYSTEMS 3ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10ML SYRINGE, 10 USP UNITS/ML; PREFILLED HEPARIN FLUSH SYRINGE
|
Back to Search Results |
|
| Catalog Number |
306521 |
| Health Effect - Clinical Code |
Bacterial Infection (1735)
|
| Date of Event |
03/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Four lot numbers were provided for this incident.The information for each lot number is as follows: medical device lot #: 711571n; medical device expiration date: 10/24/2018; device manufacture date: 5/1/2017.Medical device lot #: 716771n; medical device expiration date: 12/15/2018; device manufacture date: 6/21/2017.Medical device lot #: 718091n; medical device expiration date: 12/28/2018; device manufacture date: 7/8/2017.Medical device lot #: 721481n; medical device expiration date: 2/1/2019; device manufacture date: 8/8/2017.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the (b)(4) confirmed a case of serratia marcescens bloodstream infection from (b)(6) with pfge patterns indistinguishable from the original co outbreak pattern.The patient is a (b)(6) year old male diagnosed with lymphoma and is immunocompromised with an implanted port that was accessed in an outpatient infusion clinic.The patient presented to a hospital with positive serratia marcescens culture on (b)(6) 2018, was treated, and subsequently discharged.The specific treatment provided is unknown.The device implicated with this incident is a 3ml bd posiflush¿ heparin lock flush syringe, in 10ml syringe, 10 usp units/ml.*out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and (b)(4) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and (b)(4) identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.(b)(4)¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the cdc confirmed a case of serratia marcescens bloodstream infection from massachusetts with pfge patterns indistinguishable from the original co outbreak pattern.The patient is a 34 year old male diagnosed with lymphoma and is immunocompromised with an implanted port that was accessed in an outpatient infusion clinic.The patient presented to a hospital with positive serratia marcescens culture on (b)(6) 2018, was treated, and subsequently discharged.The specific treatment provided is unknown.The device implicated with this incident is a 3ml bd posiflush¿ heparin lock flush syringe, in 10ml syringe, 10 usp units/ml.Out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and centers for disease control and prevention (cdc) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and cdc identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.Cdc¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and cdc.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: lot numbers: 711571n, 716771n, 718091n, and 721481n for product code: 306521 were provided for evaluation by our quality engineer team.Upon reviewing the production history for the provided lot numbers, no deviations or non-conformances were identified during the manufacturing process.A corrective and preventive action plan was initiated to further investigate and monitor this issue.One-hundred and twenty retained samples for each lot number provided were visually inspected and no abnormalities were observed in regards to the solution.Forty retained samples for lot number: 721481n and 718091n were sent for sterility testing.The sterility testing confirmed that the samples exhibited no microbial growth after incubation.Multiple samples produced before and after lot numbers: 711571n and 716771n were sent for sterility testing and confirmed that no microbial growth was exhibited after incubation.This in combination with the daily environmental monitoring and sterility testing provides confidence in the sterility of the reported lot numbers.The root cause analysis of the reported infection cases under capa 350041 has not identified a direct causation between the infections and the bd franklin product.
|
| |
|
Search Alerts/Recalls
|
|
|