| Model Number |
21-FINALDRIV-TL |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history review was done and the device met all specifications.This is report 3 of 5.Several attempts have been made to get the device back for inspections and it still has not been returned.
|
| |
|
Event or Problem Description
|
|
During a slimfuse removal case the set screws stripped.The surgeon made several attempts to remove the screws and broke multiple devices.All pieces were retrieved and there was a surgical delay of 40 minutes and no patient effect.
|
| |
|