• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY EPIC¿ VASCULAR; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY EPIC¿ VASCULAR; STENT, ILIAC Back to Search Results
Model Number H74939200064020
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).  device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information from that review, a supplemental medwatch will be filed.  (b)(4).
 
Event or Problem Description
It was reported that stent placement problem occurred.The 60% stenosed target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.After a 014 choice pt guide wire crossed the lesion, a 6x40x120mm epic¿ stent was advanced for treatment.However, upon stent deployment, the stent jumped forward and missed the lesion.The stent was deployed at the wrong spot.The physician used another of the same device to complete the procedure.No patient complications were reported and patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC¿ VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7541597
Report Number2134265-2018-04471
Device Sequence Number2435467
Product Code NIO
UDI-Device Identifier08714729805281
UDI-Public08714729805281
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/22/2023
Device Model NumberH74939200064020
Device Catalogue Number39200-06402
Device Lot Number0021900923
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2018
Initial Report FDA Received Date05/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-