MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HARH36
Device Problem Melted (1385)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/13/2017
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to stryker sustainability solutions (sss) for evaluation.The submitted medwatch form did not state if the device was available for return.Multiple attempts were made to the customer to arrange for device return.However, no other information regarding device return was provided from the facility.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: use error: too much force or torque applied to instrument, or grasping/pulling use error: applying pressure between instrument blade and tissue pad without having tissue between them use error: keeping clamp arm open when back cutting or while blade is active without tissue between blade and tissue pad use error: incidental and prolonged activation against solid surfaces, such as bone, metal or plastic.The instructions for use (ifu) state: care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this occurs, there may be an instrument failure, and the generator touchscreen displays a troubleshooting message.For optimal performance and to avoid tissue sticking, clean the instrument blade, clamp arm, and distal end of the shaft throughout the procedure by activating the instrument tip in saline.Note: do not touch the instrument to metal while activated.Note: do not clean the blade tip with abrasives.It can be wiped with a moist gauze sponge to remove tissue, if necessary.If tissue is still visible in the clamp arm, use hemostats to remove residue, taking care not to actuate the hand piece.If desired, the instrument may be unplugged.Do not introduce or withdraw the instrument with the jaws open through a trocar sleeve as this may damage the instrument.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad and increased blade, clamp arm and distal shaft temperatures.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
 
Event Description
It was reported while using the harmonic scalpel, the white insulated pad at the tip started melting.Melting pieces were recovered and removed from sterile field.Multiple attempts were made to gather additional information regarding this specific event.However, no other information was provided from the facility.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer Contact
marissa richmond
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key7541723
MDR Text Key109502956
Report Number0001056128-2018-00031
Device Sequence Number1
Product Code NLQ
UDI-Device Identifier07613327346251
UDI-Public07613327346251
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161693
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/06/2018
Device Model NumberHARH36
Device Catalogue NumberHARH36RR
Device Lot Number7077227
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/27/2018
Initial Date FDA Received05/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age24 YR
-
-