MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 UROFORCE® BALLOON DILATION CATHETER; DILATION KIT WITH EAGLE INFLATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 UROFORCE® BALLOON DILATION CATHETER; DILATION KIT WITH EAGLE INFLATION DEVICE Back to Search Results
Model Number 887404
Health Effect - Clinical Codes Rupture (2208); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the uroforce catheter balloon was inserted and inflated to 4 atm using the eagle inflation device.It was reported that the physician then asked the nurse to inflate the catheter balloon to 8 atm, at which time the nurse began to use the eagle inflation device to increase the pressure in the balloon.It was reported that, when the pressure got to 5 atm, the balloon of the catheter burst.It was reported that there were no missing pieces; however, the patient's ureter "ruptured," and a stent was placed to allow the area to heal.
 
Event or Problem Description
It was reported that the uroforce catheter balloon was inserted and inflated to 4 atm using the eagle inflation device.It was reported that the physician then asked the nurse to inflate the catheter balloon to 8 atm, at which time the nurse began to use the eagle inflation device to increase the pressure in the balloon.It was reported that, when the pressure got to 5 atm, the balloon of the catheter burst.It was reported that there were no missing pieces; however, the patient's ureter "ruptured," and a stent was placed to allow the area to heal.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "device description: the uroforce® balloon dilation catheter is a dual-lumen catheter with a nydex* balloon mounted on its distal tip and with two radiopaque markers beneath the balloon that define working length.The lumen labeled ¿balloon¿ is for balloon inflation.The other lumen allows the catheter to track over a 0.038¿ (.97mm) diameter guidewire and can be used for monitoring of pressure or infusion of medication and/or contrast medium.Each balloon inflates to a stated diameter and length at a specific pressure--typically at 10 atm.The balloon is wrapped clockwise (when viewed from the proximal shaft to the distal tip) around the shaft and is protected by a profile reducing sheath that is positioned over the balloon for protection before use.A stylet is inserted in the distal (guidewire) lumen beneath the balloon to prevent kinking during shipment and storage.Indications for use: uroforce® balloon dilation catheters are recommended for dilation of the urinary tract.Contraindications: do not use the uroforce® balloon dilation catheter to treat obstruction that is due to extrinsic factors.Warnings: do not exceed the recommended maximum balloon inflation pressure that is stated on the label.Extreme caution should be used when considering dilation of irregular, non-compliant tissue or in the presence of certain calculi.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: this device should be used only by a physician who has an understanding of the clinical applications, technical principles and associated risks of balloon dilation within the urinary tract.Directions for use: catheter insertion prior to use, carefully inspect the catheter for bends, kinks, or other damage which may have occurred during shipment.Do not use the product if damage is evident.Dilation procedures may be conducted under fluoroscopic guidance with appropriate x-ray equipment or by direct vision.Endoscopically place a 0.038¿ (.97mm) diameter guidewire with the flexible end in the renal pelvis.Remove the protective balloon sheath and stylet from the catheter.Advance the catheter over the guidewire.Use the radiopaque markers located under the balloon to aid in positioning.Catheter balloon inflation: confirm proper placement of the balloon within the urinary tract.Inflate the balloon with standard inflation medium (e.G., diatrizoate meglumine injection u.S.P.60% diluted as appropriate) using a 10cc or larger syringe or inflation device.Note: do not use air or any gaseous substance as a balloon inflation medium.Note: the use of an inflation device to monitor balloon pressure is strongly recommended to determine that adequate pressure is applied and that maximum limits of the balloon are not exceeded.Caution: if loss of pressure in the balloon occurs or the balloon ruptures, immediately stop the procedure, deflate the balloon and remove carefully.Do not attempt to reinflate the balloon.If, upon inspection of the balloon, it is noted that pieces of the balloon are missing, check for and endoscopically retrieve the fragments when possible.Catheter balloon deflation and removal: deflate the balloon using a 10cc or larger syringe or inflation device.Since deflation times vary with balloon sizes and shaft lengths, check fluoroscopically to confirm deflation before attempting withdrawal.Remove the syringe or inflation device from the catheter allowing ambient pressure to relax the balloon.Gently withdraw the catheter.As the balloon starts to exit the endoscope use a smooth, gentle, steady counterclockwise motion to facilitate withdrawal.Caution: if resistance is felt when either removing the guidewire from the catheter or the catheter from the endoscope, stop and consider removing them as a single unit to prevent damage to the product or tissue trauma.Applying excessive force to the catheter can result in tip breakage or balloon separation.Warning: after use, the uroforce® balloon dilation catheter and/or accessories may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Store in a cool, dry place.Do not expose to organic solvents, temperatures in excess of 85°c or steam." the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UROFORCE® BALLOON DILATION CATHETER
Common Device Name
DILATION KIT WITH EAGLE INFLATION DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7541989
Report Number1018233-2018-01872
Device Sequence Number19300663
Product Code EZN
UDI-Device Identifier10801741052948
UDI-Public(01)10801741052948
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K993840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number887404
Device Catalogue Number887404
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/02/2018
Supplement Date Received by Manufacturer05/31/2018
Initial Report FDA Received Date05/24/2018
Supplement Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
-
-