| Model Number |
R SERIES |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed and attributed to a system interconnection flex cable that was not properly seated.The cable was replaced to correct the reported problem.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device's display showed horizontal lines and no clinical information could be seen.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|