• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. LINEAR HIP; STEM, HIP, FEMORAL, LINEAR SIZE11 STD. REDUCED NECK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. LINEAR HIP; STEM, HIP, FEMORAL, LINEAR SIZE11 STD. REDUCED NECK Back to Search Results
Model Number 425-01-011
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/09/2015
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to unknown reason.
 
Additional Manufacturer Narrative
The reason for this revision surgery was for unknown reasons.The previous surgery and the surgery detailed in this investigation occurred 5.1 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to unknown reasons.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR HIP
Common Device Name
STEM, HIP, FEMORAL, LINEAR SIZE11 STD. REDUCED NECK
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7542223
Report Number1644408-2018-00476
Device Sequence Number13927760
Product Code LPH
UDI-Device Identifier00888912077873
UDI-Public(01)00888912077873
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K991325
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/28/2014
Device Model Number425-01-011
Device Catalogue Number425-01-011
Device Lot Number53916992
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/14/2016
Supplement Date Received by Manufacturer08/23/2018
Initial Report FDA Received Date05/24/2018
Supplement Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
411-00-350, LOT 53970183; 430-03-056, LOT 54015700; 497-38-000, LOT 53993342; 499-38-009, LOT 54015640; 411-00-350, LOT 53970183; 430-03-056, LOT 54015700; 497-38-000, LOT 53993342; 499-38-009, LOT 54015640
Outcome Attributed to Adverse Event Required Intervention;
-
-