| Catalog Number |
6268.NMD06 |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/22/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation showed that there was a minor damage on the inner knife, that may have caused the leakage.Production is informed to pay attention when assembling the cutter.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre (b)(4).).Details of this activity were discussed with cdrh office of compliance ((b)(6)) during a tele-conference on (b)(6) 2017.All available information has been disclosed.The complaint is not confirmed.This complaint has now been closed.
|
| |
|
Event or Problem Description
|
|
Customer is using a ust.The unit is running in the 2x mode with a stellaris pc ( 3500 ccp ).They use tdc vitrectomies and the complaint is that air bubbles come out the cutter hole in to the eye.There is no information suggesting patient harm.
|
| |
|
Search Alerts/Recalls
|