Model Number 72202468
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/24/2018
Event Type  malfunction  
Event Description
It was reported that t1 and t2 deploy together.
 
Manufacturer Narrative
One 72202468 fast-fix 360 curved needle delivery system device received.T1, t2 and suture were not returned.The depth limiter was attached.It is creased and flared.The symptoms are indicative of difficulty with reaching the desired anchoring location.The delivery needle curve has become somewhat flattened.There is not currently obvious bow or bend although temporary flexing is a cause of unintentional ancho release.A functional test confirmed that the device delivery system did cycle and actuate.No root cause related to the manufacture of the device was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-FIX 360 CURVED NDL DELIVERY SYS
Type of Device
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7542830
MDR Text Key109285801
Report Number1219602-2018-00649
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier00885554023077
UDI-Public(01)00885554023077(17)201026(10)50693309
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/26/2020
Device Model Number72202468
Device Catalogue Number72202468
Device Lot Number50693309
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received07/19/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-