Model Number 9004
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 04/14/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed there is no similar dissection complaints associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.Based on the instructions for use (ifu), the occurrence of dissection is an inherent risk of any pta procedure.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the entire left superficial femoral artery (sfa).After treatment with the lutonix devices, the patient had a grade a dissection present.The hcp performed post-dilation with one of the already used lutonix dcb's and placed a bailout stent.The hcp deemed the medical intervention successful.The samples were discarded by the user facility and are not available for evaluation.No adverse patient effects were reported.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00132.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
MDR Report Key7542923
Report Number3006513822-2018-00131
Device Sequence Number1528610
Product Code ONU
UDI-Device Identifier00801741088759
UDI-Public(01)00801741088759
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2016
Device Model Number9004
Device Catalogue Number9020413500150
Device Lot NumberGFYK2375
Was Device Available for Evaluation? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient Weight72
-
-