| Catalog Number |
400SMTSFT0306 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the embolization coil was detached and the proximal constraint sphere was not present inside the distal detachment tip (ddt).The pusher assembly was kinked in multiple locations.Conclusions: evaluation of the first smart coil revealed it was detached and kinked in multiple locations.The embolization coil was not returned so it could not be functionally tested for advancing into a microcatheter.The root cause of the reported failure could not be determined due to the returned condition of the device.Evaluation of the second smart coil revealed a kinked and bent pusher assembly.The embolization coil had a few offset coil winds along its length.During functional testing, the embolization coil was able to be advanced through a demonstration microcatheter.Resistance was experienced when the several kinks were advanced through the demonstration microcatheter.The root cause of the reported inability to advance into a parent device could not be determined.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-01035.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils).During the procedure, the hospital staff experienced difficulty loading two smart coils into a px slim delivery microcatheter (px slim) and, therefore, the smart coils were removed.The procedure was completed using additional coils.There was no report of an adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|