| Model Number |
4108-9 |
| Medical Device Problem Codes |
Disconnection (1171); Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned cable was evaluated.During evaluation, the cable passed visual inspection; eeprom contents were verified and no problems were observed through this verification.During continuity testing, the cable failed due to a broken solder joint on the green conductor inside the mini-d connector.When connected to a monitor and a sensor, "sen off" error message displayed.The cable was confirmed to be malfunctioning.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device stopped working during active patient monitoring.The signal dropped out and would not provide a reading.No consequences or impact to patient were reported.
|
| |
|