• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 SUMMIT BASIC CEMENTED SZ 5; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 SUMMIT BASIC CEMENTED SZ 5; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 157006110
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Fibrosis (3167); Not Applicable (3189)
Date of Event 04/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Der states that this patient's hip was revised due to aseptic loosening of the summit basic cemented femoral stem at the cement to implant interface, depuy cement was used.Fiberous tissue between the stem and its surrounding cement mantel was observed by surgeon during the revision procedure.He seemed to think that the stem may have been inadvertently moved before the cement was fully hardened, creating space between the implant and the cement.The hemiarthroplasty hip was converted to a total hip.Doi: (b)(6) 2017; dor: (b)(6) 2018; right hip.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT BASIC CEMENTED SZ 5
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7544144
Report Number1818910-2018-60490
Device Sequence Number18876643
Product Code JDI
UDI-Device Identifier10603295059783
UDI-Public10603295059783
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K023453
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number157006110
Device Lot NumberD17082606
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/25/2018
Supplement Date Received by Manufacturer11/06/2018
Initial Report FDA Received Date05/25/2018
Supplement Report FDA Received Date11/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
-
-