Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The nonconformances documented for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use (ifu) states the following precaution: "during placement and use, care must be taken to avoid cutting, crimping, or damaging components." if additional pressure is applied to accessory components, this could contribute to accessory component damage.Prior to distribution, all flow 20 percutaneous endoscopic gastrostomy set - pulls are subjected to a visual inspection and functional testing to ensure device integrity.A review of the nonconformances documented for the lot number said to be involved was reviewed and confirmed that this lot met manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an isolated occurrence.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided that the fellow did not introduce the trocar correctly, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|