Model Number S7
Device Problem Material Frayed (1262)
Patient Problem No Patient Involvement (2645)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.The representative replaced the cable.The navigation system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation device being used outside a procedure.It was reported that the axiem cord was frayed and the sheath had come out showing exposed wires.There was no patient involvement.
 
Manufacturer Narrative
The power data cable for the electromagnetic interface box was returned to the manufacturer for evaluation.Testing found that the shield wire was exposed but no bare conductors were exposed.The cable was noted to pass an electrical evaluation.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7544583
MDR Text Key109278833
Report Number1723170-2018-02287
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received06/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-