MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926016270
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the moderately tortuous and severely calcified left anterior descending artery.A 2.75x16mm synergy ii stent was advanced to treat the target lesion.However, after deploying the stent, the balloon ruptured at 16 atmospheres.The device was completely removed.Post-dilatation was performed using a non-compliant balloon and the procedure was completed.No patient complications were reported and the patient's status was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: stent delivery system was returned for analysis.The stent was deployed in the patient and therefore not returned for analysis.The tip was visually and microscopically examined and no damage was noted.The balloon cones were reviewed and no issues were noted.The balloon wings appeared relaxed from their original folded position and it appeared that positive pressure was applied and a vacuum pulled.The balloon was inflated with no issues.The balloon deflated within specification as per the directions for use.The encore inflation device was verified before and after use.A visual and tactile examination of the device found multiple hypotube kinks along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the inner and outer lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the moderately tortuous and severely calcified left anterior descending artery.A 2.75x16mm synergy ii stent was advanced to treat the target lesion.However, after deploying the stent, the balloon ruptured at 16 atmospheres.The device was completely removed.Post-dilatation was performed using a non-compliant balloon and the procedure was completed.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7545009
Report Number2134265-2018-04625
Device Sequence Number13504100
Product Code NIQ
UDI-Device Identifier08714729840107
UDI-Public08714729840107
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/06/2019
Device Model NumberH7493926016270
Device Catalogue Number39260-1627
Device Lot Number21753324
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2018
Initial Date Received by Manufacturer 05/03/2018
Supplement Date Received by Manufacturer05/31/2018
Initial Report FDA Received Date05/25/2018
Supplement Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-