Catalog Number CAT8TORQ85 |
Device Problems
Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311); Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/19/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the cat8 was fractured approximately 1.5 cm from the hub.The cat8 was kinked approximately 69.5 and 94.0 cm from the hub.Conclusions: evaluation of the cat8 revealed that the device was fractured.If the device is forcefully retracted at extreme angles during removal from the packaging, damage such as this may occur.Further evaluation of the returned device revealed kinks in the catheter shaft.These kinks were likely incidental and may have occurred while packaging the device for return to penumbra.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
During preparation for a thrombectomy procedure, the hospital staff found that the indigo system aspiration catheter 8 (cat8) was torn upon removal from the packaging.The damage to the cat8 was found prior to use.Therefore, the cat8 was not used in the procedure.There is no information regarding how the procedure was completed.
|
|
Manufacturer Narrative
|
The product lot number was not provided, therefore, the manufacturing records could not be reviewed.
|
|
Search Alerts/Recalls
|