• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 8; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 8; DXE Back to Search Results
Catalog Number CAT8TORQ85
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure using an indigo system aspiration catheter 8 (cat8).During the procedure, the physician advanced the cat8 over a guidewire while preparing to advance the cat8 into a sheath in the patient; however, the physician noticed that the tip of the cat8 was pinched.The cat8 was therefore removed, and the procedure was completed using an indigo system catd aspiration catheter (catd).There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 8
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7545565
Report Number3005168196-2018-01067
Device Sequence Number1305380
Product Code DXE
UDI-Device Identifier00814548016252
UDI-Public00814548016252
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/10/2020
Device Catalogue NumberCAT8TORQ85
Device Lot NumberF79942
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2018
Initial Report FDA Received Date05/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
-
-