| Catalog Number |
18060125 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed, any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Primary procedure, right distal tibial nail implant on a right total knee patient.It was reported that patient had a distal anterior fracture of the right tibia (cause of fracture unknown), and the surgeon was performing a reduction.As reported, surgeon used "universal rod and finger tool (aka reduction spoon).Ringed up to 9/12 in the tibia.When putting finger tool distally to get reduction, something dislodged.Surgeon used slotted mallet but was not getting the device down.When pulling the device back out a piece of the finger tool broke off 10mm up from the threaded portion (threaded portion remained in universal rod), and the remainder of the finger tool remained in the tibia".Surgeon removed the remainder of the device by removing it through the fracture site distally.Surgical delay of 30-45 minutes was reported with a total tourniquet time of 2 to 2 1/2 hours.The surgeon told the rep he wants to know why this happened.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that reduction spoon t2 femur ø9 mm was alleged of k-270 (instruments - broken, deformed, worn or scratched) could be confirmed, since the device was returned for evaluation matches the alleged failure mode.The returned product shows signs of extensive usage.The threaded portion on the rear end of the product is indeed broken.The threads around the region where the fracture occurred is highly damaged.Such a damage is caused due to extensive usage leading to wear and tear of the product.Thus based on investigation, root cause can be attributed to a user related issue.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The instructions for use instructs user that only use an instrument for its intended purpose.Improper use of instruments might lead to loss in function or usage.Always treat the instrument carefully to avoid surface damage or alterations to the instrument geometry.The design of the instrument must not be modified in any way.The stryker "instructions for cleaning, sterilization, inspection and maintenance" (literature number l24002000) help to determine whether an instrument is in good condition or must be replaced due to excessive wear or obvious defects.Before the surgical procedure, ensure that all components prepared for the procedure function correctly with each other.During the procedure, repeatedly check that the connections between the instruments or between implants and instruments required for the precise positioning and fixing are secure¿ the cleaning and sterilization guide states that ¿stryker trauma & extremities typically does not specify the maximum number of uses appropriate for re-usable medical devices.The useful life of these devices depends on many factors including the method and duration of each use, and the handling between uses.¿ it also mentions that ¿for devices that may be impacted check that the device is not damaged to the extent that it malfunctions or that burrs have been produced that could damage tissues or surgical gloves.¿ and that ¿careful inspection and functional test of the device before use is the best method of determining the end of serviceable life for the medical device.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
Primary procedure, right distal tibial nail implant on a right total knee patient.It was reported that patient had a distal anterior fracture of the right tibia (cause of fracture unknown), and the surgeon was performing a reduction.As reported, surgeon used "universal rod and finger tool (aka reduction spoon).Ringed up to 9/12 in the tibia.When putting finger tool distally to get reduction, something dislodged.Surgeon used slotted mallet but was not getting the device down.When pulling the device back out a piece of the finger tool broke off 10mm up from the threaded portion (threaded portion remained in universal rod), and the remainder of the finger tool remained in the tibia".Surgeon removed the remainder of the device by removing it through the fracture site distally.Surgical delay of 30-45 minutes was reported with a total tourniquet time of 2 to 2 1/2 hours.The surgeon told the rep he wants to know why this happened.
|
| |
|