The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.One used epidural kit was received for device evaluation.Visual inspection with the unaided eye found two pieces of one catheter, one of whihc was inserted into the yellow closed epifuse connector.Under magnification, protruding wires and uneven edges were reveled on each piece of the catheter.These physical characteristics suggest that an excessive force was applied either to remove the catheter either from the patient or from the connector, which resulted in breakage.The breakage point was found to be approximately 48 cm (~19 inches) from distal tip of the catheter (patient end) when measured with calibrated ruler.The scenario of interaction with epidural needle (actual needle was not provided for the evaluation) and removal from patient could not be replicated.The piece of the catheter, inserted into the connector was released with no problem after opening the connector per device instruction for use by using a male luer slip.These instructions for use also contain warning and precautions regarding the catheter removal from the patient and excessive force application during the procedure.The catheter, epidural needle and connector components from the unopened tray were used to detect for any potential issues with the product.No issues were observed when threading the catheter through the needle, removing the needle over the catheter and during insertion and removal of the catheter from the epifuse connector.Based on the available evidence and investigation results, the reported issue was confirmed and determined to be caused during use in the clinical environment.
|