• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANATOMAGE INC. ANATOMAGE GUIDE; SURGICAL GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANATOMAGE INC. ANATOMAGE GUIDE; SURGICAL GUIDE Back to Search Results
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Code Available (3191)
Event Date 01/05/2018
Event Type  Injury  
Manufacturer Narrative
Device evaluation did not show any malfunction.Fit issue is likely caused by stone model distortion that was not detected during software qa and tissue flap requirements that was not performed prior to seating the guide on patient's mouth.
 
Event Description
Lower guide did not fit.Doctor said he had to free hand most of the case and used guide for a few sites.However, doctor was not confident in their placement.He grafted the sites and free-handed the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANATOMAGE GUIDE
Type of Device
SURGICAL GUIDE
Manufacturer (Section D)
ANATOMAGE INC.
303 almaden blvd.
suite 700
san jose CA 95110
Manufacturer (Section G)
ANATOMAGE INC.
303 almaden blvd.
suite 700
san jose CA 95110
Manufacturer Contact
meera savaliya
303 almaden blvd
suite 700
san jose, CA 95110
4088851474
MDR Report Key7546839
MDR Text Key109356558
Report Number3008272529-2018-00019
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 04/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age62 YR
-
-