Model Number LNQ11 |
Device Problems
Failure to Interrogate (1332); Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that, approximately five weeks after the implant procedure, the implantable cardiac monitor (icm) was not able to be interrogated by programmers and a remote monitor due to a suspected telemetry issue.It was also reported that the icm experienced "battery failure." the device was explanted and replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Analysis of the returned device was inconclusive.Hybrid analysis confirmed the crystal errors and depleted battery but could not determine the cause.If information is provided in the future, a supplemental report will be issued.
|
|