Model Number 51900
Medical Device Problem Codes Corroded (1131); Battery Problem (2885)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 03/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Suspected meter was evaluated by ok biotech.Error message "e-b" means battery cannot provide enough power for a test.After we placed new batteries into meter, the meter still showed e-b.We check battery compartment and found the corrosion because of battery leakage.User might exposed meter to water or high temperatures or did not use the meter for more than 3 months but keeping batteries into meter.
 
Event or Problem Description
End-user alleged that on (b)(6) 2018 around 0730 the paramedics were called due to symptoms of hypoglycemia.End-user alleged the meter would not provide a result and would display the low battery error code, e-b.It is unknown when the end-user last received a numerical result on the prodigy meter.Paramedics arrived and performed a blood glucose test on the end-user and received a 20 mg/dl with their meter.End-user was treated with iv fluids (unknown concentration of fluids) to raise his blood glucose.End-user was transported to the hospital, upon arrival end-user alleged his blood glucose was 20 mg/dl and he received more iv fluids to raise his blood glucose level.End-user remained in the hospital for 5 days and no other treatment was provided.Upon discharge the end-user's blood glucose was reported to be 180 mg/dl.End-user was advised to follow up with pcp and to start a sliding scale with basaglar.End-user stated he did not know the prescribed sliding scale.No additional details were provided in regards to this medical event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODIGY VOICE
Common Device Name
BLOOD GLUCOSE MONITORING DEVICE
Manufacturer (Section D)
OK BIOTECH C., LTD.
no. 91, sec. 2
gongdao 5th road
hsinchu, 30070
TW  30070
MDR Report Key7548144
Report Number3005862821-2018-00029
Device Sequence Number13906745
Product Code NBW
UDI-Device Identifier00384840519008
UDI-Public00384840519008
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K073118
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number51900
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2018
Initial Report FDA Received Date05/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;