| Catalog Number |
5530-G-611 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Triathlon tka procedure with a size 6 baseplate.A size 6 x 11 insert was placed on the baseplate and an attempt to secure it into the baseplate was made.It appeared to be fully seated, but when the insert was checked if locked, the surgeon was able to lift the insert out of the baseplate.A second 6 x 11 insert was opened and the surgeon was able to secure that one to the baseplate.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding seating/locking issues involving a triathlon insert was reported.The event was not confirmed by product inspection.A material analysis engineer indicated, "damages observed on insert consistent with attempted implantation.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined." a device history review confirmed all devices accepted into finished goods conformed to specification.No other events were reported for the lot indicated.The event could not be confirmed nor the root cause could be determined because the device was returned damaged.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Triathlon tka procedure with a size 6 baseplate.A size 6 x 11 insert was placed on the baseplate and an attempt to secure it into the baseplate was made.It appeared to be fully seated, but when the insert was checked if locked, the surgeon was able to lift the insert out of the baseplate.A second 6 x 11 insert was opened and the surgeon was able to secure that one to the baseplate.
|
| |
|