The reported event that distal anterolateral tibia plate axsos 3 ti for left tibia 6 hole / l127mm was alleged of 'tissue damaged' could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event, such as pictures, or x-rays as well as the affected device must be available in order to determine the root cause of the complaint event.Nevertheless, based on complaint history, some possible root causes are, but are not limited to: a potential misuse by the user.Indeed, the surgeon who performed the revision surgery stated that: "[.] the surgeon did comment that he thought if the most medial distal hole has been filled this event may have been averted.".Indeed, according to the operative technique, the axsos plates are designed in such a way that "[.] the system design is based on clinical input from an international panel of experienced surgeons, data from literature, and both practical and biomechanical testing.The anatomical shape, the fixed screw trajectory, and high surface quality take into account the current demands of clinical physicians for appropriate fixation, high fatigue strength, and minimal soft tissue damage.An improper seating of the screw in the plate, which could have damages the soft tissues surrounding it.An improper size selection / placement of the plate in the anatomic area, which could have lead to damage the soft tissues surrounding it.An extensive reading and understanding of the operative technique and instruction for use are necessary to avoid this type of events.Indeed, the ifu reminds the user that: "[.] ''ensure that you are familiar with the intended uses, indications/ contraindications, compatibility and correct handling of the implant, which are described in the operative technique manual for the product system.Please remember that pro duct systems may be subject to alterations that affect the compatibility of the implant with other implants or with instruments.For your information, avail yourself of the training courses and publications offered (e.G.Operative techniques).A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|