| Model Number |
MMT-723NAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a blank display as insulin pump was not working.The customer¿s blood glucose level 88mg/dl at the time of the incident.Customer also stated that the pump was not expose to moisture.The customer was assisted with troubleshooting and the display did not return.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Pump received with blank display due to corroded battery tube spring contact and corroded battery tube.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify unexpected battery power loss anomaly due to blank display.
|
| |
|