| Model Number |
460.039 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient¿s height is reported as 5 feet 10 inches.Patient¿s date of birth/age is unknown.Date of postoperative plate breakage is unknown.Additional device product codes: hwc, jdq.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter phone number is not provided for reporting.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested and is currently pending completion.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was implanted with ti sternal locking angled plate on an unknown date.Post-operatively on (b)(6) 2018, it was found out that the plate was broken off due to unknown reason.It was also reported that the patient underwent a surgery to remove the ti sternal locking plate on an unknown date.No further information was reported.This is report 1 of 1 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Patient ethnicity is unknown.Part: 460.039; lot: 6177979; part manufacture date: july 09, 2009; manufacturing location: elmira.Part expiration date: n/a (for s part numbers); nonconformance¿s noted: dhr record review: a review of the device history record (dhr) revealed no complaint related anomalies.The device history record shows this lot of titanium (ti) sternal locking angle 12 hole plates was processed through the normal manufacturing and inspection operations with no non-conformance reports.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.The review of the raw material device history record revealed this lot met all specifications with the above-mentioned thirteen non-conformance reports noted.These nonconformances had no relationship to the plate breakage.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|