• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493808008450
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The hypotube and shaft were examined.The mid-shaft was separated 121.5cm from the hub.The separated end was stretched and jagged indicating that the separation was due to tensile overload.The distal end of the device was not returned for analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on 02-may-2018.It was reported that crossing difficulties were encountered. the 90% stenosed, 10x4.5mm target lesion was located in the moderately tortuous and moderately calcified right coronary artery.A 4.5mm x 8mm quantum¿ maverick¿ balloon catheter was advanced; however, the device was not able to cross the target lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed the shaft detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7549947
Report Number2134265-2018-04491
Device Sequence Number233411
Product Code LOX
UDI-Device Identifier08714729392613
UDI-Public08714729392613
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2020
Device Model NumberH7493808008450
Device Catalogue Number38080-0845
Device Lot Number0021162128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2018
Initial Report FDA Received Date05/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Patient Weight65
-
-