Device is a combination product.Device evaluated by mfr: synergy ii us mr 2.50 x 8mm stent delivery system was returned for analysis.A visual examination of the crimped stent identified distal stent damage.Distal strut segment 1 and 2 were damaged with struts lifted and pulled distally.The remainder of the stent was undamaged.The undamaged crimped stent od (outer diameter) was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.Using a handheld encore inflation device, the balloon was inflated to rated burst pressure.The stent was deployed and the balloon was deflated within the specification set out in the directions for use.A visual and tactile examination of the device revealed no issues.An examination of the shaft polymer extrusion found no issues.The bi-component bond showed no signs of damage.There was hardened medium present in the shaft polymer extrusion.The tip was visually and microscopically examined and damage was noted.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|